Internal: 300w x 302d x 500h mm
Packed: 43 x 43 x 57cm
Featuring double skinned walled chamber for insulation purposes
with removable shelves for ease of loading.The lid is designed to
ensure condensed steam runs back into the chamber. Lifting handles
on both the unit and lid which features an 18mm diameter
thermometer/thermocouple entry port.
For water level control the Koch Steriliser has a filling port and
water level indication in an external reservoir. Connect laboratory
tubing to filling port and a water supply, adjust the water flow to
gently top up the water level, to maintaining boiling.
This unit is not intended for sterilisation of equipment used in
clinical care of patients.
The unit is protected by a thermal cut out which is manually reset.
This is a Traditional benchtop steam steriliser, displaces air passively from the chamber and load by steam generated within the steriliser. It is known as a ‘Type N’ cycle. Being the simplest type and least expensive to purchase, operate and maintain.
Suitable for Load type 'Non-wrapped solid items.' Not to be used for hollow devices or wrapped loads.
Drain and clean chamber and reservoir at the end of each day and leave dry. Replenish with sterile water for irrigation from an unopened container.
No accessories available for this product.
Internal: 400w x 400d x 500h mm
Packed: 610 x 540 x 551mm
Internal: 400w x 400d x 500h mm
Packed: 610 x 540 x 551mm
Featuring double skinned walled chamber for insulation purposes
with two removable shelves for ease of loading.
The lid is designed to ensure condensed steam runs back into the chamber. Lifting handles on both body and lid with an 18mm diameter thermometer/thermocouple entry port.
For stability mounted on four feet with filling/water level indicator reservoir for safety.
The unit is protected by a thermal cut out which is manually reset.
- 110 mins
This is a Traditional benchtop steam steriliser, displaces air
passively from the chamber and load by steam generated within the
steriliser. It is known as a ‘Type N’ cycle. Being the simplest
type and least expensive to purchase, operate and maintain.
Suitable for Load type 'Non-wrapped solid items.' Not to be used
for hollow devices or wrapped loads.
Drain and clean chamber and reservoir at the end of each day and
leave dry. Replenish with sterile water for irrigation from an
unopened container.
3. Features two removable shelves with looped handles making for easy retrieval and ease of loading/unloading.
4. The lid is designed to ensure condensed steam runs back into the chamber.
5. The lid has a 18mm diameter thermometer port.
No accessories available for this product.
Internal: 470w x 770d x 580h mm
Packed: 800w x 1200d x 800h mm
Featuring double skinned walled chamber for insulation purposes with removable shelves for ease of loading.The lid is designed to ensure condensed steam runs back into the chamber. Lifting handles on both the unit and lid which features an 18mm diameter thermometer/thermocouple entry port.
For water level control the Koch Steriliser has a filling port and water level indication in an external reservoir. Connect laboratory tubing to filling port and a water supply, adjust the water flow to gently top up the water level, to maintaining boiling.
This unit is not intended for sterilisation of equipment used in clinical care of patients.
The unit is protected by a thermal cut out which is manually reset.
Always read the instruction book.